Product Details:
|
COLOR: | BLACK AND BLUE | MATERIAL: | ABS |
---|---|---|---|
SIZE: | 175(L) ×110(W) ×70(D) Mm | POWER: | 5 Alkaline Battery |
WEIGHT: | 0.7KG | Packing: | 28pcs/carton, Carton Size:43.5*37.5*51.5cm |
High Light: | LCD Display Digital Insulation Tester,Digital Insulation Tester 1000V,Data Recording Megohmmeter |
Large LCD Display Digital Insulation Tester 1000V 100V 250V 500V
VICTOR 60H 100V 250V 500V 1000V Digital Insulation Tester Resistance Tester With Data Recording Megohmmeter Large LCD Display
Basic features of the product
Basic measurement functions
DC voltage DCV
Measurement range: 0 ~ ±1000V.
Measurement basic accuracy: 2%
AC true RMS voltage ACV
Measurement range: 30~750V(50/60Hz)
Measurement basic accuracy: 2%
Insulation resistance test
4 insulation test voltage ranges: 100V, 250V, 500V, 1000V
Basic accuracy: 0~+20%
Insulation test range: 0.00 MΩ ~ 19.99 GΩ
Measurement basic accuracy: 5%
Short circuit current is about 1.3 mA
Insulation testing other features
Insulation test operating modes, including:
Single, continuous test
Insulation timed test
Insulation PI (polarization index) test: the ratio of the resistance value measured when the insulation excitation voltage is applied for 10 minutes to that of 1 minute
Insulation DAR (absorption ratio) test: the ratio of the resistance value measured when the insulation excitation voltage is applied for 60 seconds to that of 15 seconds
Insulation resistance and voltage monitoring test display
Insulation resistance test automatic range
Test voltage automatic discharge function
Voltage detection and alarm of the measured object
The TEST switch with lock function
Insulation comparison test, the measurement result can be evaluated by the preset "pass/fail (✓ / ✘)" limit and be given audio alarms accordingly.
Data recording / viewing function: manual recording, internal memory can independently store 100 (groups) measurement data, recorded data can be viewed on the instrument.
Product calibration
By connecting with the standard instrument, calibrating panel according to the procedures and saving the appropriate calibration data, the periodic calibration is completed and the due accuracy and performance of the product are guaranteed.
Segment code LCD screen display, white backlight.
Easy man-machine operation, portable, solid and suitable for field use.
Product basic functions and performance (technical data)
(applicable for one year after calibration, 23℃±5℃, 0~70%RH, accuracy = ±(%reading+LSD))
Insulation resistance measurement (RISO)
Rated voltage | 100V | 250V | 500V | 1000V |
Measurement range |
0.00MΩ~19.99MΩ 20.0MΩ~199.9MΩ |
0.00MΩ~19.99MΩ 20.0MΩ~199.9MΩ 200MΩ~499MΩ |
0.00MΩ~19.99MΩ 20.0MΩ~199.9MΩ 200MΩ~999MΩ |
0.00MΩ~19.99MΩ 20.0MΩ~199.9MΩ 200MΩ~1999MΩ 2.00GΩ~19.99GΩ |
Open circuit voltage |
DC 100V + 20%,-0% |
DC 250V + 20% ,-0% |
DC 500V + 20% ,-0% |
DC 1000V + 20%,-0% |
Drop resistance | 80kΩ | 200kΩ | 450kΩ | 1MΩ |
Short circuit | 1.3 mA nominal value | |||
Accuracy | 5%+5 | |||
l Insulation test automatic discharge time: when capacitor C = 1µF or less, discharge time<1 second. l Insulation test alive circuit test: if the terminal voltage is greater than 30 V before the test is initialized, the alarm will be activated to prohibits the test. l Insulation test maximum capacitive load: operative below 1µF. |
Voltage monitoring display (insulation resistance measurement)
Measurement function | Measurement range | Resolution | Accuracy |
DC voltage DCV | 0~1000V | 1V | 2%+3 |
Voltage test
Measurement function |
DC voltage DCV |
AC RMS voltage ACV |
Measurement range | 0~±1000V |
30~750V Frequency band:45Hz~200Hz |
Resolution | 1V | |
Accuracy | 2%+3 |
General features of the product
Operating temperature and humidity:0 to 40℃ (non-condensing ≤10℃), ≤85%RH
Storage temperature and humidity:-20℃ to 60℃,≤90%RH
Working altitude:≤2000m.(Indoor use)
Rated transient overvoltage:6K (1.2*50us) I, II, III measurement peak value
Withstand voltage:AC8320V(50/60Hz)/5 seconds (between the circuit and the outer box)
Insulation impedance:1000MΩ or greater/DC1000V (between the circuit and the outer box)
Vibration and shock:Randomness 2g, 5 to 500Hz
1m drop test:Calibration period and warm-up time
To ensure the accuracy of this table, the calibration period is one year or other
The warm-up time after turned on is 10 minutes or other
Power supply:Powered by 8 x 1.5 V AA alkaline batteries
15V power adapter power supply
Automatic shutdown operation: automatic shutdown time 0~90 minutes can be set
Size and weight:178 × 110 × 59 (mm), about 600g (including batteries)